Hanshin HS-3041v

  • All processes such as heating, sterilization, vacuum drying are fully automated.
  • In case of errors included self-test feature to install six types of violations, allowing you to easily fix the problem.
  •   Autoclave ensures operator protection by: System lock doors, safety valve, auto power off, overheat protection, etc.
  • Built-in steam generator and vacuum pump.
  • This model is widely used both in
    laboratories and in small clinics.
# loop12345
  download type:An UnboxedIn the packageAn UnboxedIn the packageLiquid
Working temperature:134 ° with134 ° with121 ° with121 ° with121 ° with
Exposure time:5 min15 min20 min30 min30 min
Drying time:15 min30 min30 min30 min
Exhaust:Fast cooker hoodWithout hoods
ModelHS-2519VDHS-2321VDHS-3041VD
Overall size:WxHxD = 483h403h607 mmWxHxD = 483h387h613 mmWxHxD = 581h468h703 mm
Unit size:WxHxD = 483h362h502 mmWxHxD = 483h360h496 mmWxHxD = 581h441h556 mm
Camera:The 254х375 мм
Volume of 19.5 liters,
steel SUS 316
Cylindrical type
Square type
WxHxD = 225h225h425mm
Volume of 21.5 litres
ITS 31 steel
Square type
WxHxD = 300h300h450mm
40.5 l volume,
steel SUS 316
Water tank:WxHxD = 238h233h117mm
Volume 4 l
Input tray:WxHxD = 184h18h340mm
Quantity: 4 PCs
WxHxD = 216h44h405mm
Quantity: 2 PCs
WxHxD = 291h44h432mm
Quantity: 2 PCs
Operating temperature:121 ° with-134 ° with
Sterilizing agent:Saturated vapor pressure of 2.1 Bar, max
Office:Microprocessor
Vacuum pump:Flow rate 20 l/minWater consumption 30 l/min
Drying efficiency:Humidity, porous fabric less 2%
Air filter:The minimum filtration efficiency: 99.97% for 0.3 m µ
Pressure Gauge:Scale: 4 Bar ~ 760mmHg
Printer:Option
Power requirements:230V AC, 50/60 Hz
Power consumption:1600 W2000 W
Weight:40 kg55 kg77 kg