DiaSys Respons 920

Trusted Performance

The respons®920 is a random access bench top analyzer with a true throughput of 200 tests/hour (360 tests/hour with ISE) irrespective of whether mono- or 2-component tests are performed.

Customer-oriented

With an on-board capacity of 30 different methods in mono- or twin containers, 30 sample plus nine positions for calibrators, controls or emergency samples, respons®920 offers almost unlimited flexibility in organizing laboratory workflows. Ready-to-use reagents, adapter free one-grip loading of reagent containers, integrated bar code reader for samples and reagents as well as an intuitive software interface ensues in an allround system for everyday use.

Economic

Not only is the reagent consumption low, but DiaSys reagents are also well-known for their long-term stability. Broad measuring ranges minimize the need for retesting.

Technical Specifications

System typeBench top random access clinical chemistry analyzer
Throughput200 tests/hour with a cycle time of 18 seconds for mono- and 2 component tests<br>360 tests/hour with ISE
Sample tray30 positions for bar coded patient samples including STAT positions, 9 positions for blanks, calibrators, controls, samples without barcode and ISE solutions
Sample typesSerum, plasma, urine, CSF, whole blood
Reagent tray30 different methods in bar coded mono- or twin-containers for adapter-free one-grip loading, refrigerated
Sample volume2 – 70 μL
Reagent pipetting volumeReagent 1: 50 – 300 μL<br>Reagent 2: 10 – 200 μL
STAT-analyticsSeveral sample positions available
Bar code identificationAutomatic bar code scan for reagents and samples
Measuring principleColorimetry (Rate/End Point); Turbidimetric Immunoassay; Direct Potentiometry (ISE: K, Na, Li, Cl)
CalibrationLinear, non-linear, multi point
Sample tubes/cupsMost commonly used blood collection tubes and sample cups
Sample dilutionDilution ratio: 2 to 150 times
Reaction temperature37 ± 0.2 °C
Reaction unitTemperature controlled rotating tray with 45 reusable quartz glass cuvettes (37 ± 0.2 °C)
Photometry8 wavelengths: 340, 405, 450, 505, 546, 578, 660 and 700 nm (mono- and bichromatic)
Photometric linearity and resolutionLinearity: 0 – 2.5 OD; resolution: 0.0001 OD
Water consumptionUp to 7.5 liters per hour
System interfaceAnalyzer – PC: USB connectivity bi-directional; CPU: Pentium IV or higher
Power sourceAC 220 V ± 10 %, 50 ± 1 Hz or AC 110 V ± 10 %, 60 ± 1 Hz ; 600 VA (excluding PC/printer/monitor)
Dimensions81 cm (W) x 70 cm (D) x 60 cm (H)
WeightApproximately 110 kg